Thermo Scientific™ SPECTRONIC™ 200 Spectrophotometer Accessories

Product Code. 15390187 Shop All Thermo Scientific Products
Change view
Click to view available options
Type:
Cell Holder
Cuvette Rack
Replacement Sample Compartment
SPECTRONIC Standards 2 Kit
Unit Size:
1 set
Each
Pack of 6
4 product options available for selection
Product selection table with 4 available options. Use arrow keys to navigate and Enter or Space to select.
Product Code. Type Unit Size
15390187 SPECTRONIC Standards 2 Kit 1 set
11833784 Cuvette Rack Pack of 6
15287116 Cell Holder Each
15661794 Replacement Sample Compartment Each
Use arrow keys to navigate between rows. Press Enter or Space to select a product option. 4 options available.
4 options
This item is not returnable. View return policy
Product Code. 15390187 Supplier Thermo Scientific™ Supplier No. 840253100
  • ONLINE EXCLUSIVE PRICE
    1815.00€
    1687.50€ / 1 sets
    Ahorro 127.50€
    7% Descuento
Fecha estimada de envoi: 09-12-2026
Add to basket
Add to basket

missing translation for 'validMainframeMsg'

This item is not returnable. View return policy

Accessories for Thermo Scientific™ SPECTRONIC 200 Spectrophotometer

Set of 7 glass pieces mounted in metal frames sized to fit standard square cuvette holders.
 

  • Four calibrated neutral density glass standards for checking photometric accuracy at nominal values of 0.3A, 0.5A, 1A and 1.5A
    • Standards are calibrated at 440.0n, 465.0nm, 546.1nm, 590.0nm and 635.0nm
  • One calibrated standard for checking wavelength accuracy at nominal wavelengths on 400nm, 522nm and 777nm
    • The symmetry of the peaks means that tests of wavelength accuracy are independent of the spectral bandwidth (between 1nm and 10nm) of the instrument being tested
  • 1 filter for checking stray light in UV-Vis instruments at 220nm
  • 1 filter for checking stray light in Vis instruments at 400nm
  • Each set of SPECTRONIC Standards is individually tested and certified to be accurate, with values traceable to the National Institute of Standards and Technology (NIST) or other Primary Standards
  • Filters are supplied in a tough plastic storage container
  • Recalibration service is available from Thermo Fisher Scientific.  Recommended recalibration interval is two years


Note: Stray light filters should block light completely at the measurement wavelength and are therefore not something that can be calibrated.